categories

JP-2002G METTLER SCALE

JP-2002G METTLER SCALE Common Process for Balance Limit values ..

JP-2002G METTLER SCALE

Common Process for Balance

Limit values

Maximum capacity

2200 g

Readability

0.01 g

Repeatability (at nominal load)

0.01 g

Linearity deviation

0.02 g

Sensitivity temperature drift

3 ppm/°C

Typical values

Repeatability (at nominal load)

0.007 g

Minimum sample weight (acc. to USP)

21 g

Minimum sample weight (U=1 %, k=2)

1.4 g

Minimum sample weight OIML

0.5 g

Settling time

1 s

Internal adjustment

yes, FACT

Weighing technology

MonoBloc

Interfaces

2 ( COM1, COM2)

Balance dimensions (W x D x H)

184x290x84 mm

Weighing pan dimensions

170×190 mm

Weight of balance

3.4 kg

Weights for routin testing

Large Weight/Class OIML/ASTM

2000 g / F2/4

Small Weight/Class OIML/ASTM

100 g F2/4